Document Type : original article

Authors

1 Assistant professor, Neonatal Health Research Center, Research Institute for Children's Health, Shahid Beheshti University of Medical Sciences, Tehran, Iran

2 Division of Neonatology, Department of Pediatrics Children’s Medical Center Tehran University of Medical Sciences

3 Department of Pediatrics, Children’s Medical Center, Tehran University of Medical Sciences, Tehran, Iran

4 Maternal, Fetal, and Neonatal Research Center, Family Health Institute, Tehran University of Medical Sciences, Tehran, Iran

10.22038/ijp.2024.76501.5399

Abstract

Peripherally inserted central catheters (PICCs) have long been used to provide intravenous therapy to premature neonates. The advantages of PICCs for neonates include avoidance of the pain and handling associated with repeated peripheral intravenous cannulation as well as the ability to safely deliver concentrated parenteral fluid or nutrients. Malpositioned PICC line tips can cause life-threatening complications. There are different intra- or post-procedural techniques to confirm line-tip placement to decrease the potential complications. These include plain radiographs, fluoroscopy, ultrasonographic detection, digital imaging, computed radiography, and intra-cavitary/ trans-esophageal/ and thoracic electrocardiographic monitoring. Each method has advantages and disadvantages and some methods have benefits in specific situations. The present study provides a literature review of common methods developed for detection of PICC tip positions over the last two decades.

Keywords

  1. Sharma D, Farahbakhsh N, Tabatabaii SA. Role of ultrasound for central catheter tip localization in neonates: a review of the current evidence. The Journal of Maternal-Fetal & Neonatal Medicine. 2019; 32(14):2429-37.
  2. McCay AS, Elliott EC, Walden M. Videos in clinical medicine. PICC placement in the neonate. The New England journal of medicine. 2014; 370(11):e17-e.
  3. Bloemen A, Daniels AM, Samyn MG, Janssen RJ, Elshof J-W. Electrocardiographic-guided tip positioning technique for peripherally inserted central catheters in a Dutch teaching hospital: Feasibility and cost-effectiveness analysis in a prospective cohort study. The Journal of Vascular Access. 2018; 19(6):578-84.
  4. Noonan PJ, Hanson SJ, Simpson PM, Dasgupta M, Petersen TL. Comparison of complication rates of central venous catheters versus peripherally inserted central venous catheters in pediatric patients. Pediatric critical care medicine. 2018; 19(12):1097-105.
  5. Horattas MC, Trupiano J, Hopkins S, Pasini D, Martino C, Murty A. Changing concepts in long-term central venous access: catheter selection and cost savings. American journal of infection control. 2001; 29(1):32-40.
  6. Abedin S, Kapoor G. Peripherally inserted central venous catheters are a good option for prolonged venous access in children with cancer. Pediatric blood & cancer. 2008; 51(2):251-5.
  7. SAFDARI R, KADIVAR M, NAZARI M, MOHAMMADI M. Fuzzy Expert System to Diagnose Neonatal Peripherally Inserted Central Catheters Infection. 2017.
  8. Chopra V, Flanders SA, Saint S. The problem with peripherally inserted central catheters. Jama. 2012; 308(15):1527-8.
  9. Safdari R, Kadivar M, Nazari M, Nejat AF, Mohammadi M. Designing a Decision Support System to Diagnose Neonatal Clinical PICC Infection Using Fuzzy Logic. 2017.
  10. Racadio JM, Doellman DA, Johnson ND, Bean JA, Jacobs BR. Pediatric peripherally inserted central catheters: complication rates related to catheter tip location. Pediatrics. 2001; 107(2):e28-e.
  11. Kanin M, Young G. Incidence of thrombosis in children with tunneled central venous access devices versus peripherally inserted central catheters (PICCs). Thrombosis research. 2013; 132(5):527-30.
  12. Abood GJ, Davis KA, Esposito TJ, Luchette FA, Gamelli RL. Comparison of routine chest radiograph versus clinician judgment to determine adequate central line placement in critically ill patients. Journal of Trauma and Acute Care Surgery. 2007; 63(1):50-6.
  13. Bailey SH, Shapiro SB, Mone MC, Saffle JR, Morris SE, Barton RG. Is an immediate chest radiograph necessary after central venous catheter placement in a surgical intensive care unit? The American journal of surgery. 2000; 180(6):517-22.
  14. Natividad E, Rowe T. Simultaneous rapid saline flush to correct catheter malposition: a clinical overview. Journal of the Association for Vascular Access. 2015; 20(3):159-66.
  15. Spencer TR. Repositioning of central venous access devices using a high-flow flush technique-a clinical practice and cost review. The journal of vascular access. 2017; 18(5):419-25.
  16. Weber MD, Himebauch AS, Conlon T. Repositioning of malpositioned peripherally inserted central catheter lines with the use of intracavitary electrocardiogram: a pediatric case series. The journal of vascular access. 2020; 21(2):259-64.
  17. Beigi AA, Parvizian F, Masoudpour H. Application of intravenous electrocardiography for insertion of central veins dialysis catheters. Saudi Journal of Kidney Diseases and Transplantation. 2009; 20(5):794.
  18. Schummer W, Herrmann S, Schummer C, Funke F, Steenbeck J, Fuchs J, et al. Intra‐atrial ECG is not a reliable method for positioning left internal jugular vein catheters. British journal of anaesthesia. 2003; 91(4):481-6.
  19. Zadeh MK, Shirvani A. The role of routine chest radiography for detecting complications after central venous catheter insertion. Saudi Journal of Kidney Diseases and Transplantation. 2014; 25(5):1011.
  20. Katheria A, Fleming S, Kim J. A randomized controlled trial of ultrasound-guided peripherally inserted central catheters compared with standard radiograph in neonates. Journal of Perinatology. 2013; 33(10):791-4.
  21. DeWitt AG, Zampi JD, Donohue JE, Yu S, Lloyd TR. Fluoroscopy‐guided Umbilical Venous Catheter Placement in Infants with Congenital Heart Disease. Congenital heart disease. 2015; 10(4):317-25.
  22. Keckler SJ, Spilde TL, Ho B, Tsao K, Ostlie DJ, Holcomb III GW, et al. Chest radiograph after central line placement under fluoroscopy: utility or futility? Journal of pediatric surgery. 2008; 43(5):854-6.
  23. Park S-J, Chung HH, Lee SH, Kim J-E, Kim C, Lee SM. New formulas to predict the length of a peripherally inserted central catheter based on anteroposterior chest radiographs. The Journal of Vascular Access. 2021:11297298211001147.
  24. Jain A, McNamara PJ, Ng E, El-Khuffash A. The use of targeted neonatal echocardiography to confirm placement of peripherally inserted central catheters in neonates. American journal of perinatology. 2012; 29(2):101.
  25. Chang TC, Funaki B, Szymski GX. Are routine chest radiographs necessary after image-guided placement of internal jugular central venous access devices? AJR American journal of roentgenology. 1998; 170(2):335-7.
  26. Zhou L, Xu H, Liang J, Xu M, Yu J. Effectiveness of intracavitary electrocardiogram guidance in peripherally inserted central catheter tip placement in neonates. The Journal of perinatal & neonatal nursing. 2017; 31(4):326-31.
  27. Tian G, Chen B, Qi L, Zhu Y. Modified insertion of a peripherally inserted central catheter: taking the chest radiograph earlier. Critical care nurse. 2011; 31(2):64-9.
  28. Tauzin L, Sigur N, Joubert C, Parra J, Hassid S, Moulies ME. Echocardiography allows more accurate placement of peripherally inserted central catheters in low birthweight infants. Acta Paediatrica. 2013; 102(7):703-6.
  29. Liu G, Hou W, Zhou C, Yin Y, Lu S, Duan C, et al. Meta-analysis of intracavitary electrocardiogram guidance for peripherally inserted central catheter placement. The journal of vascular access. 2019; 20(6):577-82.
  30. Nadroo AM, Glass RB, Lin J, Green RS, Holzman IR. Changes in upper extremity position cause migration of peripherally inserted central catheters in neonates. Pediatrics. 2002; 110(1):131-6.
  31. Xiao A-q, Sun J, Zhu L-h, Liao Z-y, Shen P, Zhao L-l, et al. Effectiveness of intracavitary electrocardiogram-guided peripherally inserted central catheter tip placement in premature infants: a multicentre pre-post intervention study. European journal of pediatrics. 2020; 179(3):439-46.
  32. Baldinelli F, Capozzoli G, Pedrazzoli R, Marzano N. Evaluation of the correct position of peripherally inserted central catheters: anatomical landmark vs. electrocardiographic technique. The journal of vascular access. 2015; 16(5):394-8.
  33. Oliver G, Jones M. ECG-based PICC tip verification system: an evaluation 5 years on. British Journal of Nursing. 2016; 25(19):S4-S10.
  34. Rossetti F, Pittiruti M, Lamperti M, Graziano U, Celentano D, Capozzoli G. The intracavitary ECG method for positioning the tip of central venous access devices in pediatric patients: results of an Italian multicenter study. The journal of vascular access. 2015; 16(2):137-43.
  35. Yu T, Wu L, Yuan L, Dawson R, Li R, Qiu Z, et al. The diagnostic value of intracavitary electrocardiogram for verifying tip position of peripherally inserted central catheters in cancer patients: a retrospective multicenter study. The journal of vascular access. 2019; 20(6):636-45.
  36. Pittiruti M, La Greca A, Scoppettuolo G. The electrocardiographic method for positioning the tip of central venous catheters. The journal of vascular access. 2011; 12(4):280-91.
  37. Oliver G, Jones M. Evaluation of an electrocardiograph-based PICC tip verification system. British journal of nursing. 2013; 22(Sup9):S24-S8.
  38. Motz P, Von Arnim AVSA, Iyer RS, Chabra S, Likes M, Dighe M. Point-of-care ultrasound for peripherally inserted central catheter monitoring: a pilot study. Journal of perinatal medicine. 2019; 47(9):991-6.
  39. Stokowski G, Steele D, Wilson D. The use of ultrasound to improve practice and reduce complication rates in peripherally inserted central catheter insertions: final report of investigation. Journal of Infusion Nursing. 2009; 32(3):145-55.
  40. Nichols I, Doellman D. Pediatric peripherally inserted central catheter placement: application of ultrasound technology. Journal of Infusion Nursing. 2007; 30(6):351-6.
  41. Oleti T, Sankar MJ, Thukral A, Sreenivas V, Gupta AK, Agarwal R, et al. Does ultrasound guidance for peripherally inserted central catheter (PICC) insertion reduce the incidence of tip malposition?–a randomized trial. Journal of Perinatology. 2019; 39(1):95-101.
  42. Johnson KN, Thomas T, Grove J, Jarboe MD. Insertion of peripherally inserted central catheters in neonates less than 1.5 kg using ultrasound guidance. Pediatric surgery international. 2016; 32(11):1053-7.
  43. Nguyen J. Ultrasonography for central catheter placement in the neonatal intensive care unit—a review of utility and practicality. American journal of perinatology. 2016; 33(06):525-30.
  44. Royer T. Nurse-driven interventional technology: a cost and benefit perspective. Journal of Infusion Nursing. 2001; 24(5):326-31.
  45. Kadivar M, Mosayebi Z, Ghaemi O, Sangsari R, Saeedi M, Shariat M, et al. Ultrasound and Radiography Evaluation of the Tips of Peripherally Inserted Central Catheters in Neonates Admitted to the NICU. Iranian Journal of Pediatrics. 2020; 30(6).
  46. Nakamuta S, Nishizawa T, Matsuhashi S, Shimizu A, Uraoka T, Yamamoto M. Real-time ultrasound-guided placement of peripherally inserted central venous catheter without fluoroscopy. The journal of vascular access. 2018; 19(6):609-14.
  47. Ren X-L, Li H-L, Liu J, Chen Y-J, Wang M, Qiu R-X. Ultrasound to localize the peripherally inserted central catheter tip position in newborn infants. American journal of perinatology. 2019.
  48. Ohki Y, Tabata M, Kuwashima M, Takeuchi H, Nako Y, Morikawa A. Ultrasonographic detection of very thin percutaneous central venous catheter in neonates. Acta Paediatrica. 2000; 89(11):1381-4.
  49. Westergaard B, Classen V, Walther‐Larsen S. Peripherally inserted central catheters in infants and children–indications, techniques, complications and clinical recommendations. Acta Anaesthesiologica Scandinavica. 2013; 57(3):278-87.
  50. Bourgeois FC, Lamagna P, Chiang VW. Peripherally inserted central catheters. Pediatric emergency care. 2011; 27(6):556-61.
  51. Mack V, Nißler D, Kasikci D, Malouhi A, Aschenbach R, Teichgräber U. Magnetic tracking and electrocardiography-guided tip confirmation system versus fluoroscopy for placement of peripherally inserted central catheters: a randomized, noninferiority comparison. Cardiovascular and Interventional Radiology. 2020; 43(12):1891-7.
  52. Doniger SJ, Ishimine P, Fox JC, Kanegaye JT. Randomized controlled trial of ultrasound-guided peripheral intravenous catheter placement versus traditional techniques in difficult-access pediatric patients. Pediatric emergency care. 2009; 25(3):154-9.
  53. Pulickal A, Charlagorla P, Tume S, Chhabra M, Narula P, Nadroo A. Superiority of targeted neonatal echocardiography for umbilical venous catheter tip localization: accuracy of a clinician performance model. Journal of Perinatology. 2013; 33(12):950-3.
  54. Maki DG, Kluger DM, Crnich CJ, editors. The risk of bloodstream infection in adults with different intravascular devices: a systematic review of 200 published prospective studies. Mayo Clinic Proceedings; 2006: Elsevier.
  55. Singh S, Venkatesh H, Swamy R, Nagesh NK. Ultrasound guided confirmation of tip of peripherally inserted central catheter in neonates. Indian Pediatrics. 2020; 57(9):858-9.
  56. Telang N, Sharma D, Pratap OT, Kandraju H, Murki S. Use of real-time ultrasound for locating tip position in neonates undergoing peripherally inserted central catheter insertion: a pilot study. The Indian journal of medical research. 2017; 145(3):373.
  57. Thakur A, Kumar V, Modi M, Kler N, Garg P. Use of Point of Care Ultrasound for Confirming Central Line Tip Position in Neonates. Indian Pediatrics. 2020; 57(9):805-7.
  58. Jiao Y, Wu Y, Yang Z, editors. Application Value of Bedside Ultrasound in the Positioning of PICC Tips in Preterm Infants. E3S Web of Conferences; 2021: EDP Sciences.