Document Type : original article

Authors

Department of Paediatrics, University of Uyo Teaching hospital, Uyo, Akwa Ibom state, Nigeria

10.22038/ijp.2024.79267.5446

Abstract

Background: Blood pressure measurement in newborns using conventional auscultatory and palpatory methods is technically difficult. As a result, different blood pressure measuring devices such as oscillometric and doppler devices have been developed. However, in resource challenged environments, these devices are not widely available unlike pulse oximeters which are relatively cheap and portable. Thus, the objective of this study was to investigate the reliability of using the pulse oximeter in measuring systolic blood pressure in normal, term neonates whilst comparing it to oscillometric and doppler/aneroid blood pressure measuring devices.
Methods: This was a prospective, cross-sectional study involving 192 healthy terms, appropriate for gestational age neonates who were recruited. Their systolic blood pressures were measured simultaneously using oscillometry, doppler/anaeroid sphygmomanometer and pulse oximetry at 12-hour intervals for the first 48 hours of life.
Results: One hundred and eighty-five babies were analyzed. Mean systolic blood pressure readings using doppler with anaeroid sphygmomanometer was 60.35(SD 2.27), by Oscillometric device, it was 69.83 (SD 12.55) and by pulse oximetry 55.08 (SD 2.11). Systolic blood pressure readings using a pulse oximeter were significantly lower compared to the other methods and the degree of absolute agreement between the rating instruments was poor especially with oscillometric measurements where the level of agreement further declined with time.
Conclusion: Pulse oximetry is a better alternative to oscillometry in systolic blood pressure measurement in newborns especially in resource challenged environments but it cannot be used interchangeably or substituted for doppler anaeroid systolic blood pressure measurement

Keywords

  1. Lewis O, Stephen H. The measurement of blood pressure. J Hum Hypertens. 1994; 8(12):865–871.
  2. Dionne J.M, Bremner S.A, Baygani S.K, Batton B, Ergenekon E, Bhatt-Mehta V, et al. Method of Blood Pressure Measurement in Neonates and Infants: A Systematic Review and Analysis. The Journal of Pediatrics. 2020; 221:23-31.
  3. Muntner P, Shimbo D, Carey R.M, Charleston J.B, Gaillard T, Misra S, et al. Measurement of Blood Pressure in Humans: A Scientific Statement from the American Heart Association. Hypertension. 2019; 73(5):e35–66.
  4. Chijioke C, Okolo T, Nwadike K, Ejim E, Ekochin F, Aronu G, et al. Availability and functionality of sphygmomanometers at health care institutions in Enugu, Nigeria. Niger J Clin Pract. 2015; 18(4):544-547.
  5. Jubran A. Pulse oximetry. Crit Care. 1999; 3(2):R11–17.
  6. Sullivan, L. Power and Sample Size Determination (Internet). (cited 2023 Jun 14). Available from: https://sphweb.bumc.bu.edu/otlt/mph-modules/bs/bs704_power/bs704_power_print.html
  7. Nascimento M.C.V.A, Xavier C.C, Goulart E.M.A. Arterial blood pressure of term newborns during the first week of life. Braz J Med Biol Res. 2002; 35(8):905–911.
  8. Lim S.H, Kim S.H. Blood pressure measurements and hypertension in infants, children, and adolescents: from post mercury to mobile devices. Clin Exp Pediatr. 2022;65(2):73–80.
  9. Kim-Gau N.G. Clinical validation protocols for noninvasive blood pressure monitors and their recognition by regulatory authorities and professional organizations: rationale and considerations for a single unified protocol or standard. Blood Press Monit. 2013; 18(5):282–289.
  10. Raju R, Isac M. Pulse oximetry vs non-invasive blood pressure/oscillometry to record blood pressure in neonates: A prospective observational study. J Fam Med Pry Care. 2022;11(6):2685-2689.
  11. Langbaum M, Eyal FG. A practical and reliable method of measuring blood pressure in the neonate by pulse oximetry. J Pediatr. 1994;125(4):591–595.
  12. Vinayak R, Grover N, Behal M. Comparison of blood pressure measurement by pulse oximetry and oscillometry techniques in neonates. J. Evolution Med. Dent. Sci. 2017;6(93):6738-6743.
  13. Movius A.J, Bratton S.L, Sorensen G.K. Use of pulse oximetry for blood pressure measurement after cardiac surgery. Arch Dis Child. 1998;78(5):457–460.
  14. Luks A.M, Swenson E.R. Pulse oximetry for monitoring patients with COVID-19 at home. Potential pitfalls and practical guidance. Ann Am Thorac Soc. 2020;17:1040-1046.
  15. Zafar A, Garg S. Non-invasive blood pressure measurement in neonates: comparative analysis of different methods for reliability. 2015 DOI: 10.1186/ISRCTN36164200.